Equipment Manual
Siglent SDS1000CML Oscilloscope (PDF)
Equipment Videos
Important Notes
Triggering
Be sure the oscilloscope is appropriately triggered. If not, the waveforms will appear to dance around the screen. Refer to the oscilloscope triggering video for more information.
Scale
When directly comparing two signals, be sure the scales are equal. Rotate the large knobs on each channel to change the vertical (voltage) scale. The scale values are displayed in the lower left corner of the oscilloscope.
The image below shows two signals displayed with different scales. Channel 1 (yellow) has 2 V/division. Channel 2 (magenta) has 500 mV/division. While it appears from looking at the screen that channel 2 has a larger amplitude, that’s not the case. Channel 1 has an amplitude of 4 V (2 V/division * divisions). Channel 2 has an amplitude of 1.5 V (500 mV/division * 3 divisions).

Position
To correctly measure signal properties, be sure the signals are positioned without an offset from the oscilloscope. Press the small knobs on each channel to zero the position.
Zooming
When measuring timing parameters, be sure the signal is not zoomed out too much. The oscilloscope uses the number of pixels to measure time and voltage values. When zoomed out too much, these measurements will not be accurate.
The figure below shows a signal that is likely zoomed out too much. The signal on channel 2 corresponds to a homogeneous 1st order circuit transient response. The time constant should be ~490 us. However, the measurement shows a time constant of 600 us.

The figure below shows the same signal that’s zoomed in to a more appropriate degree. Now the time constant is measured at 520 us.
